Measurement of Glomerular Filtration Rate (GFR) in an Ectopic Pelvic Kidney by Dual Head Gamma Camera

Authors

  • Hosne Ara Begum Principal Medical Officer, Institute of Nuclear Medicine & Allied Science, Mitford, Dhaka
  • Mahbub ur Rahman Institute of Nuclear Medicine & Allied Science, Mitford, Dhaka
  • Samira Sharmin Institute of Nuclear Medicine & Allied Science, Mitford, Dhaka
  • Jesmin Ferdous Institute of Nuclear Medicine & Allied Science, Mitford, Dhaka
  • Jamiul Hossain Institute of Nuclear Medicine & Allied Science, Mitford, Dhaka

DOI:

https://doi.org/10.3329/bjnm.v20i2.37394

Keywords:

Ectopic Kidney, Glomerular Filtration Rate (GFR), DTPA renogram

Abstract

Background: DTPA renogram is an accepted method to measure glomerular filtration rate (GFR) of the kidneys. The function of an ectopic kidney varies on the basis of its size, shape, position and rotation. This may lead to variation in tissue attenuation and error in computed GFR and differential renal function (DRF) of each kidney. The objectives of this study was to assess  the changes in the GFR measurement of an ectopic kidney in a dual head gamma camera using anterior and posterior imaging process and its influence on quantification of total GFR.

Patients and Method: A Total 20 patients having one ectopic pelvic kidney and other normal positioning kidney were enrolled in the study. DTPA renogram images were acquired on a dual head gamma camera (Symbia T2) in anterior and posterior views simultaneously. Both anterior and posterior images data were used separately to compute the GFR. Three sets of total GFR of both kidneys were calculated separately. In set I, total GFR (ant) is equal to sum of both kidneys GFR in anterior imaging process, in set II total GFR (post) is equal to sum of both kidneys GFR in posterior imaging process and in set III total GFR (ectopic .ant + normal. post) is equal to sum of the GFR of normal kidney on posterior image and the GFR of ectopic kidney on anterior image. These three sets of total GFRs were compared with the patient’s eGFR measured by Cockcroft Gault formula.

Result: Mean age of the patient was 36.9 ± 14.6 years (range 18-70 years). Mean total GFR (ant)   was 89.2±11.6 ml/min, total GFR (post) was 82.9±13.4 ml/min and total GFR (ectopic .ant + normal.post) was 102.5±15.9 ml/min. Mean eGFR is 101.93±24.9ml/min. When these three sets of DTPA assisted GFR compare with eGFR the Pearson’s correlation coefficient r= 0.45, 0.55 (P<0.05) for GFR (ant) and GFR (post) respectively whereas, in case of GFR (ectopic .ant + normal.post) correlation coefficient r= 0.8 (P<0.01).

Conclusion: The GFR of ectopic kidney as calculated from the anterior data was significantly higher in comparison to the GFR calculated from the posterior data.

Bangladesh J. Nuclear Med. 20(2): 115-118, July 2017

Downloads

Download data is not yet available.
Abstract
25
PDF
36

Downloads

Published

2018-08-15

How to Cite

Begum, H. A., Rahman, M. ur, Sharmin, S., Ferdous, J., & Hossain, J. (2018). Measurement of Glomerular Filtration Rate (GFR) in an Ectopic Pelvic Kidney by Dual Head Gamma Camera. Bangladesh Journal of Nuclear Medicine, 20(2), 114–118. https://doi.org/10.3329/bjnm.v20i2.37394

Issue

Section

Original Articles